Main Technical Specifications
-
Acceleration range: ±0.5g to ±70g.
-
Acceleration measurement uncertainty:
-
Within (0.5g to 1g) range: 5×10⁻⁵ g (k=1)
-
Within (1g to 70g) range: 5×10⁻⁵ (k=1)
-
-
Rotational speed requirements
-
Speed range: 20 to 340 r/min
-
Speed measurement uncertainty: 2×10⁻⁵ (k=1)
-
Speed setting resolution: 2×10⁻⁴ r/min
-
Speed stability: <2×10⁻⁵ (1-minute average)
-
Speed repeatability: <1.5×10⁻⁶
-
-
Turntable surface and radius requirements
-
Table diameter: 0.9 m
-
Working radius: approximately 0.45 m
-
-
Azimuth angle of the mounting reference surface for the high-performance quartz inertial device
-
The settable azimuth angle shall be at least 0° or 180° to allow calibration of the high-performance quartz inertial device in both positive and negative acceleration directions.
-
Azimuth misalignment angle: θ ≤ 3′
-
-
Pitch misalignment angle of the mounting reference surface for the high-performance quartz inertial device: θ ≤ 10″
-
Angle between the main spindle rotation centerline and the vertical line: θ ≤ 20″
-
Slip ring requirements
-
Number of rings: 20 rings (using two wires combined)
-
Dynamic contact resistance variation: <0.05 Ω
-
Insulation resistance: 300 MΩ / 500 VDC
-
Operating current: 1 A
-
-
Load capacity
-
The centrifuge has a load capacity of 15 kg (including fixtures).
-
-
Fixtures
-
Used for mounting the high-performance quartz inertial device under test; four devices can be installed at one time.
-
-
Self-diagnostic function
-
The equipment shall have a self-diagnostic function.
-
-
Operating mode: Automatic mode. Speed control, dynamic radius measurement, and output measurement of the high-performance quartz inertial device are performed by a computer according to operator commands.
-
Safety requirements
-
Since the centrifuge is a large rotating equipment, safety protection design shall be implemented against various possible unexpected situations.
-
-
Equipment reliability
-
The equipment shall be capable of continuous operation for 8 hours under normal conditions for calibrating high-performance quartz inertial devices.
-
MTBF of the equipment > 2000 hours.
-
| No. | Name | Quantity | Remarks |
|---|---|---|---|
| 1 | Advantech Industrial Computer 610H & Industrial KVM + 5U | 1 | |
| 2 | Keysight 34420A 7½ Digit Multimeter | 2 | |
| 3 | Multi-channel automatic switching switch, built-in constant temperature oil bath for placing high-precision sampling resistors | 1 | |
| 4 | Precision centrifuge turntable | 1 | |
| 5 | Dedicated test software | 1 | |
| 6 | Test cabinet (27U) panel + 4U | 1 | |
| 7 | Turntable control unit | 1 | |
| 8 | Installation test fixture | 4 |