Main Technical Specifications
n
n
-
n
-
n
Acceleration range: ±0.5g to ±70g.
n
-
n
Acceleration measurement uncertainty:
n
n-
n
-
n
Within (0.5g to 1g) range: 5×10⁻⁵ g (k=1)
n
-
n
Within (1g to 70g) range: 5×10⁻⁵ (k=1)
n
n
n
n
-
n
-
n
Rotational speed requirements
n
n-
n
-
n
Speed range: 20 to 340 r/min
n
-
n
Speed measurement uncertainty: 2×10⁻⁵ (k=1)
n
-
n
Speed setting resolution: 2×10⁻⁴ r/min
n
-
n
Speed stability: <2×10⁻⁵ (1-minute average)
n
-
n
Speed repeatability: <1.5×10⁻⁶
n
n
n
n
n
n
n
-
n
-
n
Turntable surface and radius requirements
n
n-
n
-
n
Table diameter: 0.9 m
n
-
n
Working radius: approximately 0.45 m
n
n
n
n
-
n
-
n
Azimuth angle of the mounting reference surface for the high-performance quartz inertial device
n
n-
n
-
n
The settable azimuth angle shall be at least 0° or 180° to allow calibration of the high-performance quartz inertial device in both positive and negative acceleration directions.
n
-
n
Azimuth misalignment angle: θ ≤ 3′
n
n
n
n
-
n
-
n
Pitch misalignment angle of the mounting reference surface for the high-performance quartz inertial device: θ ≤ 10″
n
-
n
Angle between the main spindle rotation centerline and the vertical line: θ ≤ 20″
n
-
n
Slip ring requirements
n
n-
n
-
n
Number of rings: 20 rings (using two wires combined)
n
-
n
Dynamic contact resistance variation: <0.05 Ω
n
-
n
Insulation resistance: 300 MΩ / 500 VDC
n
-
n
Operating current: 1 A
n
n
n
n
n
n
-
n
-
n
Load capacity
n
n-
n
-
n
The centrifuge has a load capacity of 15 kg (including fixtures).
n
n
n
-
n
-
n
Fixtures
n
n-
n
-
n
Used for mounting the high-performance quartz inertial device under test; four devices can be installed at one time.
n
n
n
-
n
-
n
Self-diagnostic function
n
n-
n
-
n
The equipment shall have a self-diagnostic function.
n
n
n
-
n
-
n
Operating mode: Automatic mode. Speed control, dynamic radius measurement, and output measurement of the high-performance quartz inertial device are performed by a computer according to operator commands.
n
-
n
Safety requirements
n
n-
n
-
n
Since the centrifuge is a large rotating equipment, safety protection design shall be implemented against various possible unexpected situations.
n
n
n
-
n
-
n
Equipment reliability
n
n-
n
-
n
The equipment shall be capable of continuous operation for 8 hours under normal conditions for calibrating high-performance quartz inertial devices.
n
-
n
MTBF of the equipment > 2000 hours.
n
n
n
n
-
n
n
n
n
n
n
n
n
n
n
n
n
n
n
n
n
| No. | Name | Quantity | Remarks |
|---|---|---|---|
| 1 | Advantech Industrial Computer 610H & Industrial KVM + 5U | 1 | |
| 2 | Keysight 34420A 7½ Digit Multimeter | 2 | |
| 3 | Multi-channel automatic switching switch, built-in constant temperature oil bath for placing high-precision sampling resistors | 1 | |
| 4 | Precision centrifuge turntable | 1 | |
| 5 | Dedicated test software | 1 | |
| 6 | Test cabinet (27U) panel + 4U | 1 | |
| 7 | Turntable control unit | 1 | |
| 8 | Installation test fixture | 4 |